Investigational Device Exemption in Device Trials and Medicare Coverage

An investigational device exemption (IDE) is the permission, granted under 21 CFR Part 812 in the United States, that lets a medical device without marketing approval be shipped and used in a clinical study. The study it permits is usually the source of the safety and effectiveness data that the FDA later reviews in a premarket approval (PMA) application. For health economists it matters because it shapes how much evidence exists when payers and health technology assessment bodies first meet a device, and because the FDA category given to the device shapes whether Medicare pays for the device and routine care in the study.

What the exemption permits and why it exists

The regulation states its purpose as encouraging the discovery and development of useful devices, to the extent consistent with public health, safety and ethical standards. An approved IDE allows a device that would otherwise need premarket approval or compliance with a performance standard to be shipped lawfully for an investigation. It exempts the device from a list of duties that apply to marketed devices, including registration, listing, premarket notification, premarket approval and most quality system requirements.

The FDA explains that clinical studies of devices are most often run to support a PMA, while only a small share of premarket notification [510(k)] submissions need clinical data. Unless a study is exempt, it must have an approved IDE before it starts.

Significant risk, non-significant risk and exempt studies

FDA guidance describes three types of device study under the regulation, which are significant risk, non-significant risk and exempt studies. The type sets how much regulatory work comes before the first patient is enrolled, and so how quickly evidence can start to accumulate.

A significant risk device is one that presents a potential for serious risk to the health, safety or welfare of a subject, for example an implant, a life-supporting device or one of substantial importance in diagnosing or treating disease. For these studies the sponsor must submit an IDE application to the FDA, and both the FDA and an institutional review board (IRB) must approve before the investigation begins. An investigation may begin 30 days after the FDA receives the application unless the FDA tells the sponsor it may not, or once the FDA approves the IDE by order.

Study typeFDA application neededMain requirements
Significant riskYes, approved before startAll of Part 812, including progress reports to the FDA
Non-significant riskNo, treated as approvedAbbreviated requirements, with the IRB acting for the FDA
ExemptNoPart 812 does not apply, apart from investigator disqualification

A non-significant risk study is any study that does not meet that definition. It follows abbreviated requirements, including IRB approval, informed consent and monitoring, and is treated as having an approved IDE. The sponsor makes the first risk determination, the IRB reviews it, and the FDA is the final arbiter. FDA guidance notes that such a study can start as soon as the IRB approves it. Exempt investigations include some non-invasive diagnostic devices and custom devices.

One ground on which the FDA may refuse or withdraw approval is that the risks to subjects are not outweighed by the expected benefits to them and the importance of the knowledge to be gained. The test weighs harm to participants against the social value of information.

Charging, early access and who pays for device trials

Part 812 limits how far a sponsor can recover the cost of a device trial before the device is marketed. A sponsor or investigator may not promote or test market an investigational device before FDA approval for commercial distribution. Nor may it charge subjects or investigators more than the price needed to recover the costs of manufacture, research, development and handling.

A treatment IDE allows a device under study to be used outside the trial for patients with a serious or immediately life-threatening disease and no comparable or satisfactory alternative, while the sponsor actively pursues marketing approval. It gives early access and further safety and effectiveness data before general marketing.

CMS records that the Medicare Prescription Drug, Improvement, and Modernization Act of 2003 allowed Medicare to pay routine costs of care for beneficiaries in certain categories of IDE study, removing a financial barrier to taking part.

Category A and Category B devices in Medicare coverage

When the FDA approves an IDE it places the device in one of two categories and tells CMS at the same time as the sponsor. A Category A (experimental) device is one for which the absolute risk of the device type has not been established and the FDA is unsure whether the device type can be safe and effective. A Category B (nonexperimental or investigational) device is one for which the incremental risk is the main question, or whose device type is known to be capable of being safe and effective, for example because other makers have approval or clearance.

The category sets what Medicare can pay for. For a Category A study that CMS finds meets its coverage criteria, Medicare covers routine care items and services, but not the device itself, which CMS describes as statutorily excluded. For a Category B study, Medicare may pay for the device and for routine care, provided that CMS decides before the first related claim that the study meets its coverage criteria. Payment for a Category B device may not exceed the amount Medicare would have paid for a currently used device that serves the same medical purpose and has been approved or cleared. Routine care means covered items and services that would have been given even outside the study.

The coverage criteria in 42 CFR 405.212 read in part like an appraisal of research value. The study must aim to test whether the device improves health outcomes, must not unjustifiably duplicate existing knowledge, and must have a design and sample size adequate to answer its question. It must also be registered on ClinicalTrials.gov, plan the release of results on all pre-specified outcomes including negative ones, and explain how far the results generalise to Medicare beneficiaries.

The FDA's 2017 guidance on categorisation allows a device to move from Category A to Category B for later studies once new data resolve the initial questions of safety and effectiveness.

Worked example of trial cost shares under each category

The figures below are hypothetical and chosen for illustration. An IDE study enrols 200 Medicare beneficiaries. Each receives a device costing USD 30,000 to supply, routine care costing USD 12,000 and study-specific research costing USD 8,000. A marketed device serving the same purpose is paid at USD 18,000. Medicare is assumed to pay routine care at cost, the sponsor bears everything else, and cost sharing and bundled hospital payment are ignored.

CS=n (d+r+s)−n (a r+b p)C_S = n\,(d + r + s) - n\,(a\,r + b\,p)

Here CSC_S is the sponsor's cost for Medicare participants, nn is the number of Medicare participants, dd is the device cost, rr is routine care cost, ss is research cost, pp is the Medicare device payment, and aa and bb equal 1 when Medicare covers routine care or the device and 0 otherwise.

Step 1 (total cost). Per participant c=30,000+12,000+8,000=50,000c = 30{,}000 + 12{,}000 + 8{,}000 = 50{,}000 USD, so the study costs T=200×50,000=10,000,000T = 200 \times 50{,}000 = 10{,}000{,}000 USD whoever pays. Without CMS approval (a=b=0a = b = 0) the sponsor bears all of it.

Step 2 (Category A). Medicare pays routine care of MA=200×12,000=2,400,000M_A = 200 \times 12{,}000 = 2{,}400{,}000 USD, so CS=10,000,000−2,400,000=7,600,000C_S = 10{,}000{,}000 - 2{,}400{,}000 = 7{,}600{,}000 USD.

Step 3 (Category B). Medicare also pays D=200×18,000=3,600,000D = 200 \times 18{,}000 = 3{,}600{,}000 USD for devices, so MB=2,400,000+3,600,000=6,000,000M_B = 2{,}400{,}000 + 3{,}600{,}000 = 6{,}000{,}000 USD and CS=10,000,000−6,000,000=4,000,000C_S = 10{,}000{,}000 - 6{,}000{,}000 = 4{,}000{,}000 USD.

ItemNo CMS approvalCategory ACategory B
Medicare pays (USD)02,400,0006,000,000
Sponsor pays (USD)10,000,0007,600,0004,000,000
Medicare share0%24%60%

The total cost is fixed and only its distribution changes. Category B status moves 60% of the cost for Medicare participants to the public payer, which in effect buys part of the evidence it will later use for its own coverage decision. The payment cap leaves the gap of g=30,000−18,000=12,000g = 30{,}000 - 18{,}000 = 12{,}000 USD per device with the sponsor.

How IDE evidence feeds economic evaluation and HTA

Because IDE studies usually supply the clinical data for a PMA, their design limits any later cost-effectiveness analysis. Dhruva and colleagues reviewed 78 PMAs for high-risk cardiovascular devices approved between 2000 and 2007. Only 33 of 123 supporting studies (27%) were randomised, 17 (14%) were blinded, 51 of 78 PMAs (65%) rested on a single study, and 187 of 213 primary end points (88%) were surrogate measures.

Of 193 high-risk therapeutic devices approved through the PMA route from 2014 to 2023, Mooghali and colleagues found that the main premarket study had several arms with concurrent controls for 95, a single arm with nonconcurrent controls for 89 and a single arm with no controls for 9. By July 2025, 68 of the 193 devices (35.2%) had had at least one serious recall. Single-arm designs with historical controls give no direct estimate of the incremental effect against the relevant comparator, and surrogate endpoints must be linked to final outcomes before they can yield QALYs.

CMS guidance notes that Medicare beneficiaries have historically been underrepresented in the studies supporting marketing authorisation, because of age, access, disability, comorbidity and concurrent treatment, and the Medicare generalisability criterion asks IDE sponsors to address this at the design stage. Tarricone and colleagues describe device features that challenge current assessment methods, including learning curve effects, incremental innovation that complicates the choice of comparator, and organisational impact with dynamic pricing. A value of information analysis or an early health technology assessment at the planning stage can show whether a larger or randomised design is worth its cost.

Boundaries with marketing approval, IND and coverage with evidence development

An IDE permits a device to be studied and makes no finding that it works. The regulation forbids representing an investigational device as safe or effective for the purposes under study. Marketing needs a separate FDA decision, such as premarket approval or 510(k) clearance. The investigational new drug application (IND) plays the same role for medicines under 21 CFR Part 312.

Coverage with evidence development (CED) is a different Medicare tool, in which a national coverage determination covers an item or service only within approved clinical studies or with further data collection. IDE coverage applies only to devices under an approved IDE, whereas CED can apply to any item or service whose evidence is not yet adequate for coverage under the reasonable and necessary standard. Both resemble the managed entry agreements used elsewhere, in which a payer links access to further evidence.

Limitations and common misreadings

The FDA can refuse an IDE for a scientifically unsound investigation, but an approved IDE says nothing about whether the design will answer the questions a payer or HTA body asks. It is also specific to United States law.

  1. An IDE means the device has been approved. It allows lawful shipment for study only, and safety and effectiveness remain to be shown.
  2. A non-significant risk study escapes oversight. The FDA need not be told about the study, but the IRB acts for the FDA and the abbreviated requirements still apply.
  3. Medicare pays for the device in every IDE study. Medicare never pays for a Category A device, and a Category B device is covered only when CMS accepts the study, at no more than the payment for the existing device.
  4. IDE evidence is ready for economic evaluation. Single-arm designs, surrogate end points and limited Medicare-age enrolment often leave large uncertainty in the incremental effect.
  5. The category is fixed for the life of the device. The FDA can move a device from Category A to Category B for later studies when new data resolve the initial questions.

Sources

Institutional Perspectives (2)

  • FDA

    Approved IDE Needed Before a Device Clinical Study Starts

    The FDA states that an investigational device exemption allows an investigational device to be used in a clinical study to collect safety and effectiveness data, and that such studies are most often conducted to support a PMA. Unless exempt, every clinical evaluation of an investigational device needs an approved IDE before the study starts. Requirements include an investigational plan approved by an institutional review board, informed consent from all patients, investigational-use labelling, monitoring and required records and reports, with FDA approval of the IDE as well when the study involves a significant risk device.

    FDA, Investigational Device Exemption (IDE) web page, content current as of 22 August 2024View source →
  • Centers for Medicare & Medicaid Services

    Medicare Coverage in Category A and Category B IDE Studies

    CMS states that the Medicare Prescription Drug, Improvement, and Modernization Act of 2003 allowed Medicare to pay the routine costs of care for beneficiaries in certain categories of IDE studies, removing a financial barrier to taking part. Since 1 January 2015 CMS has reviewed and approved IDE studies centrally against added coverage criteria. Approval of a Category A (Experimental) study allows coverage of routine care but not of the device, which is statutorily excluded, while approval of a Category B study allows coverage of the device and routine care.

    CMS, Medicare Coverage Related to Investigational Device Exemption (IDE) Studies web page, page last modified 10 March 2026, describing 42 CFR Part 405 Subpart BView source →

Library

Publications

16
  • Other

    Medicare Coverage Related to Investigational Device Exemption (IDE) Studies — Centers for Medicare & Medicaid Services, 2026 (Centers for Medicare & Medicaid Services)

    Source listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Journal article

    Strength of study evidence examined by the FDA in premarket approval of cardiovascular devices — Dhruva SS, Bero LA, Redberg RF, Vol. 302, No. 24, pp. 2679-2685 ed., 2009 (JAMA)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture articles on Premarket Approval and Investigational Device Exemption, where it is cited as a reference for that concept.

  • Journal article

    Does mode matter? A comparison of telephone, mail, and in-person treatments in contingent valuation surveys — Maguire KB, Vol. 90, No. 11, pp. 3528-3533 ed., 2009 (Journal of Environmental Management)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Interviewer Bias, where it is cited as a reference for that concept.

  • Journal article

    The iterative use of economic evaluation as part of the process of health technology assessment — Sculpher M, Drummond M, Buxton M, Vol. 2, No. 1, pp. 26-30 ed., 1997 (Journal of Health Services Research and Policy)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Moving Target Problem, where it is cited as a reference for that concept.

  • Book

    The development and strengthening of methods for valuing EQ-5D-5L, an overview — Stolk E, Ramos-Goñi JM, Ludwig K, Oppe M, Norman R, Chapter 2 ed., 2022 (Value Sets for EQ-5D-5L: A Compendium, Comparative Review and User Guide (Devlin N, Roudijk B, Ludwig K, eds))

    Book or book chapter listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Interviewer Bias, where it is cited as a reference for that concept.

  • Journal article

    Claims, errors, and compensation payments in medical malpractice litigation — Studdert DM, Mello MM, Gawande AA, et al, Vol. 354, No. 19, pp. 2024-2033 ed., 2006 (New England Journal of Medicine)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Medical Malpractice, where it is cited as a reference for that concept.

  • Journal article

    Controlling symptoms versus minimizing negative treatment impact: a grounded theory study of heterogeneity in patient preferences for psoriasis treatments — Szabo SM, Hawkins NS, Germeni E, Vol. 20, Article 604216 ed., 2026 (Patient Preference and Adherence)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Grounded Theory, where it is cited as a reference for that concept.

  • Journal article

    Health technology assessment capacity development in low- and middle-income countries: experiences from the international units of HITAP and NICE — Tantivess S, Chalkidou K, Tritasavit N, Teerawattananon Y, Vol. 6, Article 2119 ed., 2017 (F1000Research)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Knowledge Brokering, where it is cited as a reference for that concept.

  • Journal article

    What type of clinical evidence is needed to assess medical devices? — Tarricone R, Boscolo PR, Armeni P, Vol. 25, No. 141, pp. 259-265 ed., 2016 (European Respiratory Review)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Moving Target Problem, where it is cited as a reference for that concept.

  • Journal article

    Premarket clinical evidence strength and recalls of high-risk therapeutic medical devices — Mooghali M, Arshad F, Rathi VK, et al, Vol. 7, No. 8, Article e262626 ed., 2026 (JAMA Health Forum)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture articles on Premarket Approval and Investigational Device Exemption, where it is cited as a reference for that concept.

  • Journal article

    Challenges in the assessment of medical devices: the MedtecHTA project — Tarricone R, Torbica A, Drummond M, Vol. 26, No. Suppl 1, pp. 5-12 ed., 2017 (Health Economics)

    Journal article listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Other

    21 CFR Part 812, Investigational Device Exemptions (sections 812.1, 812.2, 812.3, 812.7, 812.20, 812.30, 812.36, 812.42, 812.66) — US Food and Drug Administration, eCFR as of 1 October 2026 ed., 2026 (Electronic Code of Federal Regulations)

    Source listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Other

    42 CFR Part 405 Subpart B, Medical Services Coverage Decisions That Relate to Health Care Technology (sections 405.201 to 405.212) — Centers for Medicare & Medicaid Services, eCFR as of 1 October 2026 ed., 2026 (Electronic Code of Federal Regulations)

    Source listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Guidance

    FDA Categorization of Investigational Device Exemption (IDE) Devices to Assist the Centers for Medicare and Medicaid Services (CMS) with Coverage Decisions — US Food and Drug Administration, 2017 (US Food and Drug Administration)

    Guidance document listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Other

    Investigational Device Exemption (IDE) — US Food and Drug Administration, 2024 (US Food and Drug Administration)

    Source listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

  • Guidance

    Significant Risk and Nonsignificant Risk Medical Device Studies — US Food and Drug Administration, 2006 (US Food and Drug Administration)

    Guidance document listed in the Sources of the HealthEconomics.wiki Concept Architecture article on Investigational Device Exemption, where it is cited as a reference for that concept.

Frequently Asked Questions (6)

  • What is an investigational device exemption?

    An investigational device exemption (IDE) is a US FDA permit to test an unapproved medical device in a clinical study, often to support premarket approval.

    Source: FDA 2024

  • How do significant risk and non-significant risk studies differ under an investigational device exemption?

    FDA guidance describes three types of device study, which are significant risk, non-significant risk and exempt studies. A significant risk device presents a potential for serious risk to the health, safety or welfare of a subject, as with many implants and life-supporting devices. Such studies must follow all of 21 CFR Part 812 and need an IDE application approved by the FDA before they proceed. A non-significant risk study follows abbreviated requirements, including institutional review board (IRB) approval, informed consent and monitoring, and can start once the IRB approves it. The sponsor makes the first risk determination, and the FDA is the final arbiter.

    Source: FDA 2006

  • How does Medicare pay for care in an investigational device exemption study?

    CMS records that the Medicare Prescription Drug, Improvement, and Modernization Act of 2003 allowed Medicare to pay the routine costs of care for beneficiaries in certain categories of IDE study, removing a financial barrier to taking part. Approval of a Category A (experimental) study allows coverage of routine care items and services, but not of the Category A device, which is statutorily excluded. Approval of a Category B (nonexperimental or investigational) study allows coverage of both the device and routine care. Since 1 January 2015 CMS has reviewed IDE studies centrally rather than through local contractors. The category the FDA assigns therefore shifts how much of a device trial's cost falls on the sponsor.

    Source: CMS 2026

  • What criteria must an investigational device exemption study meet for Medicare coverage?

    CMS lists criteria that read in part like an appraisal of research value. The principal purpose of the study must be to test whether the device improves health outcomes of appropriately selected patients, and its results must not be expected to unjustifiably duplicate existing knowledge. The design must be methodologically appropriate, with enough participants to answer the research question confidently. The study must be registered on ClinicalTrials.gov, and its protocol must describe how results on all pre-specified outcomes, including negative ones, will be released. The protocol must also explain how far the results are expected to generalise to the Medicare population, including people eligible because of age or disability.

    Source: CMS 2026

  • Can a sponsor charge for a device used under an investigational device exemption?

    Under 21 CFR Part 812, a sponsor or investigator may not promote or test market an investigational device before the FDA approves it for commercial distribution. Nor may it commercialise the device by charging subjects or investigators more than the price needed to recover the costs of manufacture, research, development and handling. A treatment IDE allows a device under study to be used outside the trial for patients with a serious or immediately life-threatening condition and no comparable or satisfactory alternative, while the sponsor actively pursues marketing approval. The regulation also forbids representing an investigational device as safe or effective for the purposes under study.

    Source: FDA 2026 (21 CFR Part 812)

  • How does evidence from investigational device exemption studies limit economic evaluation of high-risk devices?

    Studies run under an investigational device exemption usually supply the clinical data for premarket approval (PMA), so their design limits any later cost-effectiveness analysis. Mooghali and colleagues reviewed 193 high-risk therapeutic devices approved through the PMA pathway from 2014 to 2023. The main premarket study had multiple arms with concurrent controls for 95 devices, a single arm with nonconcurrent controls for 89 and a single arm with no controls for 9. By 31 July 2025, 68 of the devices (35.2%) had been subject to at least one serious recall. Single-arm designs give no direct estimate of the incremental effect against the relevant comparator.

    Source: Mooghali et al. 2026

Trust Record

Verified by Dr Darrin Baines

British health economist

Professional identity: darrinbaines.org

Verification date: 14 Oct 2026

Content version: 1.0.1

Canonical Identity

Term code
HS-HP-HSR-068

Stable URI · Machine-readable · Resolvable · CC BY 4.0